Background: The metrological traceability of values for the catalytic concentration of several enzymes assigned to a calibration material has been assured by following the recently published International Standard ISO 18153.
Methods: A traceable value with a measurement uncertainty was assigned for the catalytic concentration of alanine aminotransferase, creatine kinase, ?-glutamyltransferase and lactate dehydrogenase in two materials from different sources...
Keywords: commercial calibrator, commutability, enzymes, metrological traceability, primary calibrator
03/2006 | Clinical Chemical Laboratory Medicine, Walter de Gruyter