European Concerted Action on Anticoagulation (ECAA): International Normalized Ratio Variability of CoaguChek and TAS Point-of-Care Testing Whole Blood Prothrombin Time Monitors

The object was to assess the variability in displayed International Normalised Ratio (INR) between monitors of the same manufacture using whole blood samples from the same subjects. Two brands of monitor, CoaguChek Mini and the TAS PT-NC were tested. 14 instruments of each brand were tested on the same day at the same laboratory by the same operator using identical blood samples to avoid between-centre differences in samples and operator technique. Whole blood samples from two normal donors and four coumarin-treated patients were tested to assess between-instrument variability of INR. Results have been coded. There was a much wider dispersion of INR on Brand B than on Brand A. One Brand A instrument failed to give a result with one of the two whole blood samples from one patient. One Brand B monitor gave an aberrant result with one of the samples from a normal subject. On both brands of monitor, INR variability appeared to be due mainly to duplication differences rather than between-instrument variability on both normal and coumarin whole blood samples.

Thrombosis and Haemostasis, Schattauer

Print ISSN: 0340-6245
Volume: 88, 12/2002
Pages: 992 - 995

Show full article (external site)

Show all available items of this journal