We evaluated a new ready to use liquid assay for the homogeneous determination of HDL-cholesterol (HDL-C; Merck, Darmstadt, Germany) in comparison to phosphotungstic acid precipitation and a homogeneous assay, based on sulfated ?-cyclodextrin and polyethylene glycol-modified enzymes (Roche Diagnostics/Boehringer Mannheim, Germany). The new liquid homogeneous HDL-C assay had inter-assay coefficients' of variation of less than 2.1 %. The method is linear up to at least 3.11 mmol/l HDL-C, but even at 4.40 mmol/l the deviation from the expected value is less than 5%.
Spinking experiments with low density lipoproteins and very low density lipoproteins proved that the new assay was specific for high density lipoproteins up to cholesterol associated with low density lipoproteins (LDL-C) and very low density lipoproteins (VLDL)-triglyceride concentrations of 18.13 and 22.60 mmol/l, respectively. Free fatty acids above 2 mmol/l did not interfere. Icteric samples with bilirubin concentrations between 170 and 400
Print ISSN: 1434-6621
Volume: 37, 05/1999
Pages: 537 - 543