Science.Online
Publisher and Institutes
Akademie Verlag
Deutsches Institut für Urbanistik
Oldenbourg Wissenschaftsverlag
Walter de Gruyter
Schattauer
You are here: Home :: Area NEM :: Engineering :: Machine construction
 
Aline Maalouf

A Validated Model for the Electromagnetic Flowmeter´s Measuring Cell

Keywords: electrode/electrolyte interface, measurement cell, impedance spectrometry, impedance spectroscopy

This paper proposes as well a measurement model for the electromagnetic flowmeter’s measuring cell that constitutes an off-line measurement model for the electromagnetic flowmeter with which it is possible to try different signal processing algorithms to estimate the flowrate value. This model was validated using impedance spectroscopy and spectrometry. Knowing the value of the electrode/electrolyte impedance it is possible as well to determine the value of the input impedance of the preamplifier that reduces the error caused by the impedance of the electrode/electrolyte interface. The value of this latter is used on the other hand to determine the value of the lower frequency limit above which the reading of the induced voltage level of the electromagnetic flowmeter can be carried out with a satisfactory signal-to-noise ratio.

tm – Technisches Messen, Oldenbourg Wissenschaftsverlag

Print ISSN: 0171-8096
Volume: 73, 06/2006
Pages: 350 - 363

Show full article (external site)

Show all available items of this journal